Disposal
The analyzer has been designed for a long and fault-free life and
has been validated for 9,999 tests. As long as the Optics Check
Cartridge is giving acceptable results, the Analyzer may be used
with confidence. Excessive wear on the mechanisms will result in
error conditions, and may indicate that the Analyzer has reached
the end of its useful life. The Analyzer must then be disposed of
according to local regulations concerning the disposal of electronic
items in general.
The Cholestech GDX Analyzer has been preprogrammed to shut
down after 9,999 tests have been performed. From test ID 9900
onwards the ID number will flash momentarily to remind the user
that the analyzer is approaching the end of its programmed life.
Shutdown Information
15SHUTDOWN INFORMATION / DISPOSAL
Cholestech GDX Analyzer User Manual
14 SAFETY / LIMITATIONS OF USE
Cholestech GDX Analyzer User Manual
Limitations of Use
• For professional use only.
• For
in vitro
diagnostic use only.
• Do not use any part of the test components beyond the marked
expiry date.
• Store the test components according to the instructions on the box.
• Always wear gloves when handling blood samples.
• In patients where the lifespan of the red blood cell is significantly
shortened, e.g. hemolytic anemia, new A1C reference ranges
should be established.
• The test solutions (first and second tube) contain 0.1% sodium
azide as preservative, which is very toxic by inhalation, in contact
with skin and if swallowed. After contact with skin, wash
immediately with plenty of water.
• Cholestech is not responsible for any injury or damage
caused by the use of the Cholestech GDX Analyzer for
purposes other than for which it is intended or by modifications
of the Cholestech GDX Analyzer.
Following testing, the instrument may be left on ready for the next
test or switched off at the main socket. If unused for a period it
should be placed in its carton to protect the optical window.
Safety
The Cholestech GDX Analyzer requires a 12-volt (±5%) DC supply
regulated at greater than 250 milliamps. This is provided. There are
no user-serviceable parts.
If the Cholestech GDX Analyzer is dirty, clean it carefully with a lint-free
cloth dampened with aqueous solutions containing 5% hypochlorite or
alcohol. Take care not to get liquid in the space for the cartridge. Do
not use more aggressive chemicals such as acetone.
The Cholestech GDX Analyzer has been fully tested according to
Electrical Safety Regulations (BS EN 61326:1998).